20/22
BEA--250127-EN-09
ELMETA
®
MDS 50
➜
Open the fastening screws on the plastic cover (M4 x 35) and re-
move plastic cover. See figure to the left. If necessary, remove the
Silikon with a knife.
➜
Put the plastic cover aside.
➜
Remove the plate located above the amplifier. See figure to the
left. The plate is attached with 4 screws M3 x 6.
➜
Unlock the plug of the dataline using an appropriate tool and re-
move dataline. See figure to the left.
➜
Unscrew earthing conductor from amplifier.
➜
Carefully remove ribbon line from amplifier, no jerks, to make sure
that the plug does not get damaged. Use an appropriate tool.
➜
Open fastening scsrews (M3 x 6) at amplifier and remove ampli-
fier. Make sure that no line gets damaged. See figure to the left.
➜
Replace by new amplifier. Make sure that no line gets damaged.
The amplifiermust lock into the rail when it is totally inserted (in its
end position).
➜
All the other components must now be assembled in the reverse
order than described above.
Make sure that there is no line between the plate and the printed
circuit board when fixing the plate above the amplifier.
➜
Seal plastic cover with Silikon.
Platte
Fastening screws
Ribbon line
Earthing conductor
Dataline
Amplifier
Fastening
screws
Pinted circuit board