BEA--250127-EN-09
17/22
ELMETA
®
MDS 50
36
reserve
37
switch limit
xx
30
115
R/W
Switch limit
38
integration time
factor
2.0
1.0
4.0
R/W
Factor for setting the integration time factor
39
transmitter on/off
1
0
1
R/W
Transmitter ON-OFF
0 = OFF
1 = ON
40
ready out
xx
0
1
R
Output "equipment ready"
1 = Output active
41
output 1
xx
0
1
R
Output 1 "reporting of particles"
1 = Output active
42
output 2
xx
0
1
R
Output 2 "control"
1 = Output active
43
lamp on
xx
0
1
R
Output "pilot lamp"
1 = Output active
44
reset input
xx
0
1
R
Entry "reset exit 1"
1 = Entry active
45
webspeed input
xx
0
1
R
Entry "Pulse transmitter web speed"
1 = Entry active
46
noise with
4.0
1.0
10.0
R/W
Signal filter
47
reset time
250
50
1000
R/W
Signal filter
48
actual
Parameters for test and service purposes,
to be modified only after consultation with E+L
49
DC level
64
20
255
R/W
DC signal level for automatic compensation
50
DC delta
8
4
100
R/W
DC signal window for automatia compensation
51
autocal time
6.0
4.0
10.0
R/W
Time window "automatic compensation finished"
52
autocal error time
1.5
1.0
10.0
R/W
Time window "automatic compensation start" time
53
webspeed in count
xx
0 32000
R
Pulse counter web speed count
9. Error location
instructions
The sig nal cannot be cancelled.
➜
Reduce the sensitivity.
➜
Check the bar display. Check whether the signal is triggered by
external electromagnetic alternating fields, e.g. of electro-motors,
transformers or fluorescent lamps. If this is the case, mount the
probe head at least two meters away from the former. If neces-
sary, select another mounting location.
➜
Check the grounding strips. They must be located on clean, rust-
free metal on the grounded machine frame.