EKO INSTRUMENTS CO., LTD. VIP Checker HC-121 Instruction Manual V2
Pg. 20
8.
When measurements are completed, the results, the measurement outputs are displayed in the
sample boxes for each Sensor Unit.
Repeat the measurements by changing the combinations so that the boxes are all filled. Once all
and R^2 are calculated and displayed.
The measurement result data up to this point are not saved yet; if the software is closed
without saving, the measurement results are lost. Always click Save button to save with
an appropriate file name, as often as possible.
In case the computer shuts down due to some
malfunction, the save data can be retrieved. (See step 11. Discontinuing and restarting calibration
process)
Determination Coefficient (R^2)
ximation value and the
is
higher as it is closer to 1, which means it is a good value as calibration.
9.
Select the measurement method, Linear Approximation or Exponential Approximation by clicking
the radio buttons provided on the Main Window.
Figure 6-21. Window Completed All Measurement
Figure 6-22. Select Linear Approximation or Exponential Approximation