Troubleshooting
Hemo
Point
®
H2
User Manual
57
Display texts are not shown in your language.
Language is set
incorrectly.
Set your language
(see Section 5.3.7.1 “Setting the language”).
Time displayed is incorrect.
Time is set incorrectly. Reset the time
(see Section 5.3.3 “Setting the time”).
Date shown in the printout or data storage is incorrect or will not
be displayed as expected.
Date setting is
incorrect.
Reset the date
(see Section 5.3.2 “Setting the date”).
Date format is set
incorrectly.
Set the date format
(see Section 5.3.7.2 “Setting date format”).
Time shown in the printout or data storage is incorrect or will not
be displayed as expected.
Time is set incorrectly. Reset the time
(see Section 5.3.3 “Setting the time”).
Time format is set
incorrectly.
Set the time format
(see Section 5.3.7.3 “Setting time format”).
Incorrect units are shown.
Setting of units is
incorrect.
Set the units of measurement
(see Section 5.3.4 “Setting the units”).
Printer does not print.
Incorrect type of
printer.
Connect to a DPU-414 printer
(see Section 5.6.1 “Connecting a printer”).
Printer not “online”.
Switch the printer to online mode (see
operating instructions for the printer).
Incorrect, or incorrectly
connected, printer
cable.
Use an original printer cable from EKF
Diagnostics or check that the plug contacts are
correctly seated
(see Section 5.6.1.1 “Connecting the printer
cable”).