33
Errors/warnings
Description and potential causes
Solution
Battery error
The batteries are discharged or expired
or there is possible corrosion of the
battery contacts.
Change the batteries.
If the battery contacts are corroded,
please contact the Lactate Scout Sport
service team.
Test result too low
The test result is below the test range of
the Lactate Scout Sport;
Water used for washing may have
diluted the droplet at the puncture site.
Repeat the test with a new test strip.
Check the code configuration. Follow
the instructions for performing the test.
Dry your hands well before collecting
the sample.
Use the Lactate Scout Check
Solution to perform a functionality test.
Please contact the Lactate Scout Sport
service team if there is a persistent
error message.
Test result too high
The test result is above the test range
of the Lactate Scout Sport
Perspiration on the skin containing a
significant proportion of lactate may
have been included in the sample.
Repeat the test with a new test strips.
Check the code configuration. Follow
the instructions for performing the test.
Carefully wash perspiration away from
the intended puncture site and dry your
hands well.
Please contact the Lactate Scout Sport
service team if there is a persistent
error message.