Select the “Functionality test”
1C
sub-menu in
configuration and display mode.
Then select the concentration of your Check
Solution
18
. The measuring device will then
ask you to insert a test strip
19
. For testing
purposes, the target value will be displayed
with the measurement unit (mmol/L) in the
symbol line.
The configured code will be displayed briefly
after you have inserted the test strip. This
code must also match the code printed on the
test strip container for the functionality test as
well. The measuring device will then ask you
to fill the test strip with Check Solution
20
.
Open the Check Solution bottle.
Wipe the opening to make sure it is clean.
Dispense and discard a single drop of control
solution.