Product information about 

the Lactate Scout Test Strips 

for single, disposable use with 

Lactate Scout Sport analyzers to 

determine lactate in whole blood.   

                 Please read carefully before using.

INTENDED   USE 

 

The Lactate Scout Sport analyzer may only be 

used with Lactate Scout Test Strips. The Lactate 

Scout Sport measuring system is designed 

to measure lactate concentration in capillary 

blood in order to assess changes in physical 

performance and conditioning in healthy people 

the field of sport. 
The Lactate Scout Sport measuring system is 

not designed for taking quantitative lactate 

measurements to support clinical diagnoses 

in medical applications. No diagnosis or 

medical action should be taken on the basis 

of the values. 
The Lactate Scout Sport measurement system is 

designed for use outside the body only. Lactate 

Scout Test Strips are only for single use.

WARNING NOTICES 

Read the user manual of your Lactate Scout 

Sport analyzer before using. Avoid any risk of 

test strips being swallowed. Keep away from 

children. The handling of blood carries the risk 

of infections. Please ensure items containing or 

contacted by blood (such as test strips, lancets 

and tissues) are disposed of safely and ensure 

that all items and devices that have been in 

contact with blood are disinfected properly. 

STORAGE

 

Do not expose test strips to moisture and direct 

sun light. Keep test strips  refrigerated. Store 

the test strips in the original container only and 

within the specified temperature range of -18 

°C / 0 °F to +8 °C / 46 °F. The expiration date 

of the test strips on the label of the test strip 

container must not be exceeded. Once the test 

strip vial is opened, the test strips are stable 

for 90 days or until the specified expiry date 

(whichever is sooner). Without refrigeration, 

store test strips  

 25 °C / 77 °F for a maximum 

of 30 days. Transport and store the test strips 

in the original vial. Take out the required test 

strips for immediate use only and close the vial 

to keep the other test strips protected. Before 

using refrigerated test strips, keep the closed 

vial at ambient temperature for at least 20 min. 

Before using test strips from deep freezer, keep 

the closed vial at ambient temperature for at 

least 2 hours. During use, the Lactate Scout 

analyzer and test strips should be at the same 

temperature.

en

Rev.01 01/2022 

INTERFERENCES 

 

Impurities such as sweat, alcohol or disinfectant 

on the skin may cause problems with the sample 

uptake or cause incorrect results. Ensure that 

the sample site is cleaned carefully with water 

and dried before puncturing the finger. Use 

only fresh blood samples for measurement. The 

intake of drugs such as paracetamol or diseases, 

infections, diets, carbohydrate-rich food or stress 

can influence individual test results. 

HANDLING (FOR SINGLE, DISPOSABLE USE) 

 

Insert a new, unused Lactate Scout Test Strip 

into the analyzer. Hold the test strip in the 

middle with clean, dry fingers and do not touch 

the tip. Upon insertion of the test strip, the 

two-digit test strip code will be displayed briefly. 

Ensure that this code is identical to the code 

printed on the test strip vial. Only use strips that 

are within their expiry date. Thoroughly clean 

and dry the intended sampling site. Puncture 

the sampling site by using a sterile safety lancet. 

Blot away the first drop of blood and use the 

second drop for the measurement. Lower the 

analyzer with the tip of the inserted test strip 

to the fresh drop of blood. Do not touch the 

skin. The required sample volume (0.2 

μ

l) will 

be drawn into the test strip automatically and 

the measurement will commence. The result will 

be displayed after 10 seconds. By removing the 

test strip, the result will be stored automatically 

(including date, time and other data) and 

the analyzer will switch off. Error messages: 

Measurement results outside the measuring 

range (<0.5 mmol/L and >25.0 mmol/L) are 

marked. In case of faults or technical problems, 

error messages appear. For detailed descriptions 

of the indications and error messages, refer to 

the Lactate Scout Sport user manual.  

 

Health note: 

There are no known health risks 

when using Lactate Scout Test Strips according 

to the instructions. If discomfort is felt after use, 

a physician should be contacted immediately. 

Persons prone to excessive bleeding or infection, 

should consult a doctor before puncturing skin 

and sampling.

TECHNICAL DATA 

 

Measurement method: Enzymatic-amperometric 

detection of lactate Reagent per test strip: 

Lactate oxidase, electron mediator, additives 

Precision: Hct range 35 – 50%: 0.5 – 6.7 mmol/L 

blood lactate 

 0.2 mmol/L,  

6.8 – 25.0 mmol/L blood lactate 

 3% /  

At hematocrit values out of the specified 

range, higher deviations are possible.  Sample 

volume: 0.2

 μ

Measuring range:

 0.5 - 25.0 mmol/L  

-18°C

+8°C

Summary of Contents for Lactate Scout Sport

Page 1: ...Lactate Scout Sport Instruction manual...

Page 2: ......

Page 3: ...written consent of SensLab GmbH Part No 7023 9018 0127 Rev 01 01 2022 Lactate Scout Sport instruction manual Manufacturer SensLab GmbH Bautzner Str 67 04347 Leipzig Germany Sales and service EKF diag...

Page 4: ...batteries 10 1 4 Switching the device on and off 11 1 5 Display 11 1 6 Code setting for the test strips 12 2 Performing a test 13 2 1 Preparing for a test 13 2 2 Sample collection and test 14 2 3 Chec...

Page 5: ...aying stored values 26 3 5 Battery level indicator 27 4 Performing a step test 28 4 1 Step test mode 28 4 2 Displaying the post exertion period 30 5 Storage cleaning and disposal 31 6 Error messages a...

Page 6: ...rips are only for single use Safety instructions PLEASE READ THE INSTRUCTION MANUAL BEFORE USE AND KEEP IT IN A SAFE PLACE Failure to comply with any text marked as a WARNING can result in serious dam...

Page 7: ...ecified in the instruction manual ATTENTION Make sure you only use accessories that have been provided or recommended by the manufacturer The measuring device must not be used if it is not functioning...

Page 8: ...menus sub menus and menu items for adjusting values e g time and date OK button 1D for enabling menus sub menus and menu items for confirming choices for activating and deactivating configuration mod...

Page 9: ...batteries When inserting the new batteries ensure that the polarity is correct Push the battery cover back in until it completely locks into place 4 If the time to replace the batteries exceeds 30 se...

Page 10: ...sed for two minutes in configuration and display mode or if the device is not used for two minutes while the test strips is inserted in the measuring device 1 5 Display An information line will be dis...

Page 11: ...n The code setting menu will open The symbol line will display the CODE symbol 7 Enter the two digits of the code starting from the left Enter the digits of the code using the two arrow keys and confi...

Page 12: ...The display will briefly show the code which is currently configured A blinking drop symbol will then be displayed to indicate that the measuring device is ready to take a measurement The ambient tem...

Page 13: ...tainer immediately after removing the test strip Never leave the container open 2 2 Sample collection and testing Wash your fingers or earlobes with clean water at the puncture site Dry the puncture s...

Page 14: ...ustic signal will sound and the test will begin A circular symbol with a progress bar will be displayed 5 After ten seconds a second acoustic signal will sound and the test result will be displayed in...

Page 15: ...e code is displayed briefly when you switch measuring mode on 7 If more than 24 tests are performed with the same code a reminder to check the code will be displayed for three seconds 8 Pressing the O...

Page 16: ...nctions and displays Settings can be changed in configuration and display mode In order to go to configuration and display mode there must not be a test strip inserted in the measuring device The conf...

Page 17: ...on 2C 1B 2 2A 2B 2C 2A 3A 3B 3C 3 3 1 1 Volume control If you choose the symbol 2A in the Device settings menu and confirm your selection by pressing the OK button you will be taken to the volume cont...

Page 18: ...nd day using one of the two arrow keys and confirm your selection by pressing the OK button Once you have confirmed your selection you will then have the option to set the time A clock will appear in...

Page 19: ...20 2A 1 2 3 4 5...

Page 20: ...device firmware and device components information If you use the arrow keys to browse through the pages in this menu item then the following information will be displayed Serial number of the measurin...

Page 21: ...unctionality test is used to check the correct functioning of the Lactate Scout Sport measuring system If there are any doubts regarding the accuracy of the test result or the correct functioning of t...

Page 22: ...layed with the measurement unit mmol L in the symbol line The configured code will be displayed briefly after you have inserted the test strip This code must also match the code printed on the test st...

Page 23: ...You can track the progress of the test on the display 22 and the lactate concentration will be displayed If the result is within the permissible range a tick will be shown in the centre of the display...

Page 24: ...ice will switch back to the main menu 25 26 1E If you want to stop the stopwatch during the 10 minute interval you can do so by briefly pressing the OK button An acoustic signal will sound several tim...

Page 25: ...7B The date and time of this measurement will be shown in the bottom section of the display 27C If the measurement was taken as part of a step or endurance test where applicable the display will also...

Page 26: ...29 There are three different display options available Batteries are full 29A Batteries are partially discharged 29B Batteries are almost fully discharged and batteries need to be changed 29C If the b...

Page 27: ...port along with the relevant configured parameters For a simple step test protocol see https www ekfdiagnostics com lactate testing for athletes html The general procedure for step test monitoring is...

Page 28: ...1st phase of the exertion phase and take the sample as described in section 2 6 Remove test strip when result is displayed 7 Repeat steps 5 to 6 during the exertion phase of the step test 5 Note that...

Page 29: ...the post exertion phase the amount of time since the last test in the exertion phase is also displayed by the measuring device in measuring mode 6 The time display is updated every 10 seconds The post...

Page 30: ...d battery housing Any sample residue Check Solution dust or other contaminants on the housing must be cleaned using a soft lint free cloth or paper towel To do this moisten the cloth with a mild clean...

Page 31: ...Error when filling the test strip Sample too small or has run contact time of test strips with sample too short interrupted sample absorption or test strip pressed against skin Repeat the test with a...

Page 32: ...ns for performing the test Dry your hands well before collecting the sample Use the Lactate Scout Check Solution to perform a functionality test Please contact the Lactate Scout Sport service team if...

Page 33: ...ctionality test with new materials after checking the code Follow the instructions for performing the functionality test Make sure that the target concentration you have selected in the menu matches t...

Page 34: ...isplay Time displayed with black background The date and time were not set when the device was configured reset Set the date and time Device fails to turn on Device fails to turn on Device fails to tu...

Page 35: ...range of the measuring device Change the batteries Please contact the EKF technical support team Defective test results Values seem too high too low For example low values after exertion or high valu...

Page 36: ...ion Slow update of the screen content Slow update of the screen content The Lactate Scout Sport uses an e paper display Please note that it may take some time for the display to update depending on th...

Page 37: ...Precision 0 5 6 7 mmol L blood lactate 0 2 mmol L 6 8 25 0 mmol L blood lactate 3 At hematocrit values out of the specified range higher deviations are possible Measuring principle Enzymatic amperome...

Page 38: ...st allocations Power supply 2 x 3 V CR2450 3V lithium batteries button cell batteries Device dimensions 91 mm L x 46 mm W x 21 mm H Weight 60 g including batteries This product contains the Roboto 201...

Page 39: ...documentation and on packaging Symbol Meaning Manufacturer Serial number Attention consult accompanying documentation observe safety instructions Biological risks Follow instructions for use Designed...

Page 40: ...41 Symbol Meaning Test strip number Do not expose to direct solar radiation Protect against liquid Direct current Used batteries must be disposed of in accordance with local laws and regulations...

Page 41: ...rt measuring device with 2 batteries CR2450 7023 0522 Lactate Scout Test Strips 24 test strips in a vial 7023 3431 Lactate Scout Check Solution Duo Pack 2 levels 1 x 4 5 5 6 mmol L and 1 x 8 9 11 1 mm...

Page 42: ...e for at least 20 min Before using test strips from deep freezer keep the closed vial at ambient temperature for at least 2 hours During use the Lactate Scout analyzer and test strips should be at the...

Page 43: ...tions please refer to the Functionality test section of the user manual 1 Display of the measurement result and evaluation check mark or error message Ensure that two digit test strip codes are entere...

Page 44: ...7023 9018 0127_EN 1 0_LSSport_Manual...

Reviews: