16
8
7
9
2.4 Checking the ambient temperature
To ensure consistent and reliable test
results are achieved under different climatic
conditions, the measuring device is equipped
with a temperature sensor. The temperature
sensor is located underneath the opening for
the test strips
9
.
2.3 Checking the code setting
The code is displayed briefly when you switch
measuring mode on
7
. If more than 24
tests are performed with the same code, a
reminder to check the code will be displayed
for three seconds
8
. Pressing the OK button
within three seconds will take you to the code
configuration menu.
WARNING
Do not touch or cover the temperature sensor
with your hand. Failure to do so results in
the risk that the temperature sensor will read
the ambient temperature incorrectly and
the measuring device may record incorrect
test results.