8. Fault finding
Table 5
Fault finding
Conditions
on page 43
The pump cannot reach stated vacuum
on page 43
on page 43
on page 44
on page 44
Pump does not maintain vacuum after power-off and pump leaks oil
on page 44
The expected process vacuum level is not reached
on page 45
Fault
The pump does not run
Cause
No voltage
Remedy
Provide power supply.
Cause
Thermal switch has tripped
Remedy
Identify cause and reset switch.
Cause
Room temperature too low
Remedy
Restore temperature to allowed value.
Cause
Motor damaged
Remedy
Contact service department.
Fault
The pump cannot reach stated vacuum
Cause
Low oil level
Remedy
Top up oil.
Cause
Oil contaminated
Remedy
Replace oil.
Cause
Pump gaskets damaged
Remedy
Contact service department.
Cause
Discharge clogged
Remedy
Check couplings and outlet.
Fault
Pump is noisy
Cause
Exhaust filter element clogged
Remedy
Replace.
Cause
Bearings damaged
07/2020 - Page 43
A35104880_B - Fault finding