Remedy
Contact service department.
Cause
Motor coupling damaged (if applicable)
Remedy
Contact service department.
Cause
Vanes worn out
Remedy
Contact service department.
Cause
Solid particles in the oil
Remedy
Change oil.
Fault
Pump runs hot
Cause
Wrong oil type
Remedy
Replace oil.
Cause
Poor room ventilation
Remedy
Install an auxiliary fan.
Cause
Fan defect
Remedy
Contact service department.
Cause
Wrong power supply to motor
Remedy
Check.
Cause
Discharge clogged
Remedy
Check couplings at outlet.
Fault
High oil consumption
Cause
High working pressure (close to atmospheric pressure)
Remedy
Check oil level frequently.
Cause
Pump runs hot
Remedy
Refer to
on page 44.
Cause
Exhaust filter element clogged
Remedy
Replace.
Fault
Pump does not maintain vacuum after power-off and pump leaks oil
Cause
Check valve damage
Remedy
Contact service department.
Cause
Tank screws or plugs loose
Remedy
Tighten.
Cause
Tank gaskets damaged
Remedy
Contact service department.
Page 44 - 07/2020
A35104880_B - Fault finding