Technical data
2.7
Tracer gas analysis
Tracer gas fugitive emission testing was carried out in accordance with the method given in
A 50 mm duct was connected to the extraction port on top of the enclosure.
Table 10 - Tracer gas test parameters
Test Parameters
Tracer gas
SF
6
(Sulphur Hexafluoride)
Tracer gas concentration
100%
Tracer gas release rate
0.25 to 1 slm
Tracer gas release points
Tracer gas flow evenly split between 2 release points
1) exhaust flange on pump
2) exhaust elbow joint within exhaust extraction cover
Table 11 - Tracer gas test system parameters
System Parameters
iXL600/
iXL1000
iXL600N/
iXL1000N
Units
Extraction flow rate:
From port on top of enclosure
180
180
m
3
/h
Volume of enclosure
0.11
0.11
m
3
Free volume of enclosure
0.069
0.068
m
3
Air changes per minute
43.5
44.3
per min
Table 12 - Worst case test results
Process Gas
Maximum
Gas Flow
(slm)
TLV/LEL
(ppm)
25% TLV/
LEL
(ppm)
SF
6
Release
rate (slm)
Max SF
6
Detected
outside
enclosure
ERC
(ppm)
Pass/
Fail
*
*
Where Pass indicates acceptable enclosure (satisfies SEMI S2 criteria of less than 25.0% of the TLV).
Chlorine
Cl
2
0.2
0.5
0.125
1
0.25
0.05
Pass
Ammonia
NH
3
20
25
6.25
1
0.25
5.00
Pass
Carbon Monoxide
CO
0.3
1
0.25
1
0.25
0.08
Pass
Fluorine
F
2
0.5
25
6.25
1
0.25
0.13
Pass
Hydrogen
H
2
30
4000
1000
1
0.25
7.50
Pass
Methane
CH
4
0.05
1000
250
1
0.25
0.01
Pass
Nitrous Oxide
N
2
O
20
50
12.5
1
0.25
5.00
Pass
Nitrogen Trifluoride
NF
3
0.5
10
2.5
1
0.25
0.13
Pass
Hydrogen Chloride
HCI
0.4
5
1.25
1
0.25
0.10
Pass
Sulphur
Hexafluoride
SF
6
0.2
1000
250
1
0.25
0.05
Pass
Silicone
Tetrachloride
SiCl4
0.1
1
0.25
1
0.25
0.03
Pass