Technical data
A540-55-880 Issue D
Contact Edwards for more information about configuration requirements for earth leakage reduction.
Short circuit current rating
(when installed with class T or
class J fuses)
200
kA
Second protective earth (ground)
conductor
Must be fitted with cross-sectional area at least equal to phase
conductor size of 6 mm
2
Typical earth leakage
*
for 200 - 230 V supply
<5
mA
for 380 - 460 V supply
<10
mA
*
Typical earth leakage values were measured at steady-state conditions.
Note that higher leakage currents may occur:
i) under transient conditions such as power on or pump acceleration or
ii) with abnormal supply configurations such as a missing or earthed phase or unbalanced supply voltages.
Table 8 - Electrical connections
Description
Mating Connector description /
external supply rating
Internal supply rating
Mains connection
Refer to installation section for
wiring diagram
Weidmuller Axial screw connector:
Insert type HDC S4 BAS
p/n:1789980000
Hood type HDC 16D TOLU 1PG21G
p/n:1652680000
Alternative: Harting Axial screw
connector
Insert type Han 40A Axial module
p/n:09140022701 x 2 off used with
p/n:09140060313 (Hinged Frame)
Hood type PG21 Top Entry
p/n:09300060442
6 mm
2
stranded wire, 13-18 mm
cable OD
PDT Interface (front)
XLR type 5-way plug
24 V d.c. 0.2 A
System/PDT Interface (rear)
XLR type 5-way plug
24 V d.c. 0.75 A
*
Ethernet interface
Standard RJ45 type or Neutrik
®
EtherCon
®
RJ45
(IEEE802.3i
10 Base T Ethernet)
EMS interface
External emergency stop switch
Pin 1 - supply, Pin 2 - return
Note:
If there is no external
connection a link plug must be fitted
to operate the pump.
XLR type 6-way plug
24 V d.c. 100 mA
Comms 24 V supply
Pin 5 - supply, Pin 6 - 0 V supply
common
24 V d.c. 0.75 A
Table 7 - General electrical data (continued)
Description
Rating
Units