Technical data
A540-55-880 Issue D
Note:
The temperature management system is a valved system and the ‘Rated Min Pressure Differential’ may not
correlate with water flow rates stated in the table under all operating conditions. The ‘Rated Min Pressure
Differential’ is required to maintain adequate cooling water flow under adverse operating conditions.
Hardness
<100
ppm of CaCO
3
(<100 mg of CaCO
3
per litre)
Resistivity
1k
≤
ρ
≤
1000k
ohm-cm
Materials in contact with cooling
water
Stainless steel, PVDF, Nitrile, PTFE
and fluoroelastomer
Water inlet connection
3/8 inch male quick connector
Water outlet connection
3/8 inch female quick connector
*
Water consumption varies with pump operating temperature and water temperature; these figures measured
at factory default internal pump temperature, 15 °C water inlet temperature and ultimate inlet pressure.
Table 9 - Water cooling system data (continued)
Description
Rating
Units