CADENCE II Fetal Monitor User Manual
- 57 -
Working System
Continuous running equipment
Earth Leakage Current (Limit)
N.C. S.F.C.
500µA 1000µA
Enclosure Leakage Current (Limit)
N.C. S.F.C.
100µA 500µA
Patient Leakage Current (Limit)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
Patient Auxiliary Current (Limit)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
LCD Size
5.7”
Active Area:
103.0(W) x 79.0(H) mm (320 x 240 dots)
Resolution:
85 dots per inch
Display
Backlight:
Cold cathode fluorescent
Signal
Interface
DB9 network interface
Ultrasound Transducer
Weight:
190g
Cable Length:
2.5m
Dimension:
88mm×35mm
TOCO Transducer
Weight:
180g
Cable Length:
2.5m
Dimension:
88mm×35mm
Remote Event Marker
Length:
2.5m
Weight:
56g
A1.3 Performance Specifications
Technique: Ultrasound
Pulse
Doppler with autocorrelation
Pulse Repetition Rate:
2 KHz
Pulse Duration:
92 µs
Ultrasound
Ultrasound Frequency: (1.0±10%) MHz