
CADENCE II Fetal Monitor User Manual
- 30 -
6.5 Adjusting the Volume
The monitor automatically detects which channel the ultrasound transducer is connected to. If
both channels are connected with US transducers, the fetal heart sound of US1 is heard by default.
You can press the
CHANNEL
key to switch the sound to US2.
Adjust the default monitoring volume:
The fetal heart sound volume returns to a default level in new monitoring after the
AUTO
key is
pressed. This default level is adjustable. To change this level,
1 Press the
SETUP
key six times to open the
Auto Monitor
menu.
2 Press the
UP
key or
DOWN
key to move the cursor to
Volume
; press the
CHANNEL
key.
3 Press the
UP
key or
DOWN
key to choose the volume from 0 ~ 9; the default is 3.
4 Press the
CHANNEL
key to confirm selection.
5 Press the
AUTO
key to exit.
Adjust the real-time monitoring volume:
If the default volume is not satisfactory during monitoring, you can press the
UP
key or
DOWN
key to increase or decrease the volume of the current channel. Press the
CHANNEL
key and
adjust the volume of the other channel in the same way.
Adjust the key volume:
When a key is pressed, the monitor gives a “Di” sound. This volume is also adjustable.
1 Press the
SETUP
key three times to open the
Fetal Monitor Configuration
menu.
2 Press the
UP
key or
DOWN
key to move the cursor to
Key Volume
; press the
CHANNEL
key.
3 Press the
UP
key or
DOWN
key to choose the key volume from
High
,
Low
(default) and
Off
.
4 Press the
CHANNEL
key to confirm selection.
5 Press the
AUTO
key to exit.
NOTE:
When the monitor does not execute the function of a key, it gives a “Do” sound.