CADENCE II Fetal Monitor User Manual
- 53 -
(1) Cleaning of Transducers and Leads
To clean the US transducer, TOCO transducer, IUP cables, and DECG lead, follow these steps:
1)
Wipe them with a soft cloth dampened in cleaning solution;
2)
Clean them with a soft cloth dampened in water;
3)
Air-dry them or wipe off the remaining moisture with a soft dry cloth.
The recommended cleansers for accessories are listed below:
Accessory Cleansers
Ultrasound Transducer
TOCO Transducer
BURATON LIQUID
MIKROZID
ETHANOL 70%
SPORACIDIN
CIDEX
DECG Leads, IUP cable, IUP
connecting cable
Mild alcohol-free soap water
CAUTION
:
1) Be sure the temperature of cleaning solutions does not exceed 45
ºC (
113
ºF
)
.
2) Do not immerse them in any liquid.
3) Only clean the outer surface of the connectors, make sure no liquid goes into the
connector.
4) After cleaning, no remaining cleanser is allowed on the surface.
(2) Cleaning of Belt
Wash soiled belts with soap and water. The water temperature must not exceed 60
ºC (140 ºF).
8.3 Disinfecting
Clean the equipment before disinfecting.
The table below lists the allowed disinfectant bases:
Type Base
Instrument Disinfectant
Glutaraldehyde up to 3.6%
Surface Disinfectant Ethanol
1- and 2- Propanol
CAUTION
:
1) Do not use any disinfectant containing additional active ingredients other than those
listed.