CADENCE II Fetal Monitor User Manual
- 8 -
2.3
Connecting the Power Cable
A
Use the power cord provided with the monitor. Put the connector of the power cord into the
power socket on the monitor. Put the plug of the power cord into a grounded 3-slot power
output special for hospital usage.
A
Make sure the AC power supply complies with the following specification: 100V-240V~,
50Hz/60 Hz.
WARNING
:
If the protective grounding (protective earth) system is doubtful, the monitor must be
powered only by the inner power.
2.4 Feeding Paper
Feeding Paper
If the paper is used up or a paper jam happens, you have to feed paper. The operation procedure is
as follows:
1)
Press the
position simultaneously on both sides of the recorder cover to open the paper
compartment.
2)
Take out the Z-fold thermosensitive paper from the wrapper. Place the pack in the
compartment with the pane facing down and the green safety area on the left.
3)
Feed the recorder paper into the paper advance slot (Figure 2-1). The paper will come out
from the notch automatically. Adjust the paper length by rotating the gear on the left of the
handle if required.
Figure 2-1 Diagram of feeding paper
4)
If the paper is slantwise, pull the handle up and rotate the gear to force the paper out. Push the
handle down and then feed paper again.
5)
After closing the cover, make sure that the paper can come out from the paper notch.
NOTES:
1) When feeding paper, the black handle must be down. If a paper jam happens, pull up
the handle first, and then push the gear to force the paper out.
2) Feed the paper gently and avoid touching the print head.
Paper advance slot
The side without print
scale faces up.
Handle