![EDAN Acclarix AX4 User Manual Download Page 145](http://html1.mh-extra.com/html/edan/acclarix-ax4/acclarix-ax4_user-manual_2357138145.webp)
Acclarix AX4 Diagnostic Ultrasound System User Manual
EMC Information
- 138 -
14 EMC Information
Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
The system is intended for use in the electromagnetic environment specified below. The customer
or the user of the system should assure that they are used in such an environment.
Emission test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The system uses RF energy only for their internal
function. Therefore, their RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class A
The system is suitable for use in all establishments,
other than domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Complies
NOTE:
The EMISSIONS characteristics of the system make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is
normally required) the system might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or re-orienting the
equipment.