![EDAN Acclarix AX4 User Manual Download Page 101](http://html1.mh-extra.com/html/edan/acclarix-ax4/acclarix-ax4_user-manual_2357138101.webp)
Acclarix AX4 Diagnostic Ultrasound System User Manual Features
- 94 -
Figure 7-21 Turning the Knob 1 Clockwise
Step 3: Insert the needle into the round hole, as shown in
.
Figure 7-22 Insert the Needle into the Round Hole
7.5.2. Activating Needle Guide Function
To enable the needle guide function:
1.
In the B mode imaging, press
Needle
button on touch screen, A needle touch screen UI is
displayed, Press
Enable
button to active the Needle Guide function.
2.
Press
Double Line
button to switch double line and single line as the Needle guide Line
graphics.
3.
Some needle guide brackets support multiple (up to 2) angles. If the current transducer
supports such a guide then the
Angle
button appears. Pressing it cycles through the available
angles.
Figure 7-23 Needle Guide Touch Screen
NOTE:
The distance between each dot of the needle guide line indicates 0.5 cm.
Round hole