ECP2050
40 / 64
Socket for connecting Ethernet is on rear panel of the unit and has two signalization diodes:
Location
Description
LED1
Meaning: OFF = no connection; orange = 10Mbps; green =
100 Mbps.
LED2
Meaning: OFF = no activity; orange = half-duplex mode; green
= fully duplex mode.
Communication with the unit can be easily tested from a computer using internet browser by
entering address http://XXX.XXX.XXX.XXX, where XXX.XXX.XXX.XXX is actual IP address of the
unit.
Current setting can be found in
Menu/Info/Ethernet
.
6.4.1.1. DHCP/STATIC IP
The choice of
DHCP
enables automatic setting of connecting parameters (IP address,
Subnet mask, Default gateway) in network, where this function is supported. Allocation of address
can take several seconds. Actual connecting parameters can be found in Menu/Info/Ethernet.
The choice of
STATIC IP
enables manual setting of connecting parameters in next
subchapters. This choice is suitable for direct connection with PC (using a cable/switch) without
other connection to another network. It is necessary to set on the unit and also in PC suitable
connecting parameters (e.g. ECP2000: IP=192.168.91.30, MASK=255.255.255.0 and PC:
IP=192.168.91.1, MASK=255.255.255.0).
Caution:
When using static address in corporate network, always contact network
administrator in order to avoid collisions!
Default setting is automatic setting of parameters using DHCP, which enables easier first
connection. Despite that it is necessary to contact network administrator, if you are in corporate
network, in order to ensure stability of connecting parameters in time.
6.4.1.2. IP Address
IP address, used in mode
STATIC IP
. Default value is 192.168.91.30.
6.4.1.3. Subnet Mask
Subnet mask, used in mode
STATIC IP
. Default value is 255.255.255.0.
6.4.1.4. Default Gateway
Default gateway, used in mode
STATIC IP
. Default value is 0.0.0.0.
It is usually not necessary to set this item.
6.4.2. Remote Keylock
If the unit is controlled or there are collected data via serial line (RS232) or using Ethernet
interface, the LED RMT is alight. The unit switches to mode of remote control immediately after
receiving first query or demand through RS232 line or Ethernet, and leaves this mode, if it does not
receive any query or demand for a period of 3s.