3. Safety instructions
3.1 Probe information
3.1.1 Probe types and corresponding applications
The probes provided with the Imago device are the following :
- LB760P
Linear Rectal probe, 7,5 MHz (this rectal probe is used for gynaecology and reproduction
diagnosis).
- LB760 S
Linear Rectal probe, 7,5 MHz (this rectal probe is used for gynaecology and reproduction
diagnosis).
- CB360S
Convex probe, 3.5 MHz (this external probe is used for the following applications : general,
abdominal, obstetrics).
- LB740R
Linear probe, 7.5 MHz (this external probe is used for the following applications: general,
orthopedics, small parts).
- SB355M Mechanical probe, 5 MHz (this external probe is used for the following applications :
general, abdominal, small parts).
- DB355M Mechanical probe 5 MHz (this rectal probe is used for gynaecology and also abdominal,
small parts).
3.1.2 Handling of probes
Warning :
The probe is fragile and requires proper handling, care and cleaning. Transducer care includes daily inspections, cleaning and
disinfections between each animal.
Please refer to chapter 7.2 about how to perform proper care to the probe.
Warning :
Do not bend or twist the transducer cable. If the transducer housing becomes cracked or broken or if there are cuts or
openings in the cable, the electrical safety features of the transducer might be compromised.
Warning :
Do not immerse the transducer connector in any liquid. Immersing it can destroy its electrical safety features.
3.1.3 Ultrasound coupling gel
Some ultrasound coupling gels and lotions can damage the probes.
Agents containing the following chemicals are known to damage transducers:
▪
Acetone
▪
Methanol
▪
Denatured ethyl alcohol
▪
Mineral oil
▪
Iodine
▪
Any lotion or gel that contains perfume.
Warning :
Check the gel contents with your gel supplier.