Instruction Manual MM850481 ed. 03a
36
Use
the
[
S
] and [
T
] keys to move the cursor symbol to “MEASURE” option.
Press the [
ENTER
] key to select the option. Press [
ESCAPE
] (F2) to return to the previous menu.
MEASURE
TIGHTNESS TEST
GAS LEAK PROBE
AMBIENT CO
T+RH% PROBE
ESCAPE
–
NEXT
Use
the
[
S
] and [
T
] keys to move the cursor symbol to “AMBIENT CO” option.
Press the [
ENTER
] key to select the option. Press [
ESCAPE
] (F2) to return to the previous menu.
Connect the probe to the unit and wait for the following page:
AMBIENT CO
Amb. CO: 7 ppm
Internal Probe
ESCAPE
–
ZERO
NOTE: W
HEN PROBE IS CONNECTED THE INSTRUMENT BEGIN A
40
SECONDS WAITING TIME
. I
F PROBE IS CONNECTED IN
A
MBIENT
CO
MENU
,
A WAITING MESSAGE WILL BE DISPLAYED
.
Before insert the probe in working area, agitate it for same seconds in fresh and uncontaminated air. Press the [
ZERO
] key to
reset the measure and start CO monitoring.
From this moment the unit will compare the monitored value with the CO limit (default OSHA limit: 35ppm). It is possible to
change the CO limit using the GasConfig software for PC.
Press [
ESCAPE
] (F2) to return to the previous menu.
NOTE: I
F THE PROCEDURE ABOVE IS EXECUTED AND THE
CO
PROBE REMAIN CONNECTED
,
THE INSTRUMENT WILL MONITOR AMBIENT
CO
ALSO DURING OTHER OPERATIVE MODE
.
If alarm condition is marked, press the [
ENTER
] key to silence it.
7.11
Natural gas leak external probe
(option)
To enable natural gas monitoring press the [
MENU
] key from measuring page. The following page will be displayed:
MENU
MEASURE
ION. CURR
TEMPERATURE
GAS VELOCITY
ESCAPE
–
Use
the
[
S
] and [
T
] keys to move the cursor symbol to “MEASURE” option.
Press the [
ENTER
] key to select the option. Press [
ESCAPE
] (F2) to return to the previous menu.