Pressure decay procedure
(option)
This is an additional feature that allows you to verify the condition of part or the entire plant. The system works with the
pressure decay principle. Be sure to isolate and empty the plant section you want to verify at first !
ATTENTION
T
HE USAGE OF A PRESSURE HIGHER THEN
±300
H
P
A WILL DAMAGE THE THE PRESSURE SENSOR OF THE INSTRUMENT IN IRREVERSIBLE
WAY
,
CAUSING A PERMANENT DEFORMATION OF THE MEMBRANE
.
7.9.1
Test A (1 minute)
This optional operative mode allows operator to find pipe leakage. To set and run leak test, press the [
MENU
] key from
measuring page. The following page will be displayed:
MENU
MEASURE
ION. CURR
TEMPERATURE
GAS VELOCITY
ESCAPE
–
x
Use the [
S
] and [
T
] keys to move the cursor symbol to “MEASURE” option.
x
Press the [
ENTER
] key to select the option. Press [
ESCAPE
] (F2) to return to the previous menu.
MEASURE
TIGHTNESS TEST
GAS LEAK PROBE
AMBIENT CO
T+RH% PROBE
ESCAPE
–
NEXT