Instruction Manual MM850481 ed. 03a
27
Press
[
ESCAPE
] (F1) to return to the previous page.
Now you can proceed with the Smoke index measurement and/or the gas analysis. The memory stored value will be printed
together the analysis.
7.3.1 Burner
pressure
measurement
Using the burner pressure probe, it is possible to measure the pressure on the burner. Connect the probe to the instrument
pressure input and connect to the burner plug. Follow the procedure described in paragraph 8.4 to read the pressure.
7.4 Smoke
measurement
To avoid condensation inside the smoke pump, it should aspirate ambient air before the gas sample. This can be obtained
aspirating the air by the fireside (not in the exhaust gas connection pipe) without using the paper filter.
Introduce now the filter in the pump. The gas sample should be drawn to the center of the combustion gas flux. The correct
gas volume is sampled with 10 uniform shoots by the pump.
Compare the darkening of the filter with the smoke index table. The smoke index found does not exceed the ones defined by
the current normative.
We suggested to use a retaining cone during gas sampling.
N
OTE
: I
F YOU WONT TO PRINT THE SMOKE INDEX WITH THE ANALYSIS DATA
,
YOU SHOULD DO THE MEASUREMENT BEFORE THE FLUE
GAS ANALYSIS
.
7.4.1
Smoke index data input
Before input the data, you have to execute the smoke index measurements. You can input up to 3 values; the analyzer will
do the average calculus.
The smoke measurement requires the instrument in normal operating conditions indicating one of the measured or
calculated parameters. Press [
SMOKE
] on the keyboard. The following page will be shown:
SMOKE
Meas. 1: -
Meas. 2: - AVG
Meas. 3: - -
Oil.Der.: NO
ESCAPE
–
Press
[
S
]and/or [
T
] to move the cursor symbol to the “Measure 1” input field.
Press
[
ENTER
] to input the value. See instruction on paragraph 8.1.4.
If necessary, input the “Measure 2” and “Measure 3” smoke index. The instrument will compute the average.
Press
[
S
] and/or [
T
] to move the cursor symbol to the “Oil derivative”.
Press
[
ENTER
] to change the value for the correct fuel type.
Press
[
ESCAPE
] (
F1
) to return to the previous page without save the changes.
Press
[
ENTER
] to memory store the selection and return to the previous page.
7.5 Combustion
air
temperature
(option)