(5) Do not overlap the recording paper, otherwise the ECG record may trans-print each other;
(6) Pay attention to the size of the paper. Improper size may damage the printer head or Roller shaft.
2. Safety Information
2.1 Safety Warning
Use three core power cord and protective grounding socket when mains supply is applied, in
order not to cause electric shock to animal and operator.
Be sure that the installation room has stable power supply system which is reliable and
grounding.
When the system is not complete and reliable, cut off the A/C power and use internal D/C
supply directly
Do not use the instrument near anesthetic gases, oxygen, hydrogen or other flammable and
corrosive chemicals.
Do not use the instrument near high-voltage, high- static, X ray, ultrasonic, electrosurgical
equipment and other strong magnetic wave environment.
Only qualified service engineers can install this equipment. And only service engineers
authorized by ECGMAC can open the shell case.
Auxiliary equipments connected to the digital and analog interfaces must be certified
according to IEC standards (e.g. IEC950 for data processing equipment and IEC60601-1 for
medical equipment). Therefore anybody, who connects additional equipment to the signal input or
output connector to configure a medical system, must make sure that it complies with the
requirements of the valid version of the system standard IEC60601-1-1. If any questions, please
consult our service department or the local agent on your side.
When defibrillator is used simultaneously with the instrument, the operator should not