ECG grid print: enable and disable grid printing on the recording paper
Sample Mode
Pre sample mode
10s ECG data sampled before pressing the START/STOP key will be printed out.
Real-time Mode
10seconds ECG data sampled after pressing the START/STOP key will be printed out.
The data will be recorded as soon as you pressing the "START/STOP"button.
Trigger Mode
After pressing the START/STOP key, if arrhythmia ECG data is detected during the learning course,
the printing will be triggered automatically. Otherwise, no printout come out.
Periodic Mode, Duration and Interval
Sample mode is in periodic for the same data. For example, if Interval is set to 2 min, and Duration
is set to 24 min, after pressing the START/STOP key, the printing will be performed every two
minutes and come to 12 times.
Print time
The total print time can be adjusted from 5 to 60seconds.
Press
【
F1
】
button to save print setup and return, or press
【
F2
】
button to cancel and return.
5.5.3 Auto Analysis Setup
Press button
【
2
】
to enter auto analysis setup at the system setup interface. The interface is shown as
below: