13
Do not clean the unit and accessories with abrasive fabric and avoid scratching the electrodes.
Be sure no cleanser remains on the unit, animal cable or electrodes.
Disinfection, if required, can not be done with high temperature, autoclaving or radiation.
Do not use chloric disinfectant such as chloride and sodium hypochlorite etc.
Maintenance and repair should be applied on the main units and accessories regularly (at
least once a year)
ECG machines are classified into Measuring Instrument, so users should send it to Official
Measurement Administrative to test and certify according to national metrological calibration
regulation of electro-cardiograph and electroencephalogram machine every year.
The instrument signal input/output connector (when needed) must be connected with Class I
equipment which is IEC60601-1 compliant, and the total leakage current should be tested to be
available by users themselves.
Electrical schematic diagram and parts lists are only available to qualified repair stations or
staff of ECGMAC authorized.
2.5 Intended Use
The intended use is to acquire the ECG waveforms from animals through electrodes. It should be
used by doctors or trained nurses in medical institutions. The ECG signals recorded by the
instrument can help users to analyze and diagnose heart diseases. The multi-channel ECG with the
measured value and its explanatory notes are only for the reference of clinicians.