45
Earth leakage current,Limit: NC 5OOuA,SFC 1000 uA
Enclosure leakage current,Limit: NC 100µA, SFC 500µA
animal leakage current,Limit: 10 uA (CF Type
)
animal auxiliary current,Limit: NC a.c. 10µA,d.c. 10µA
;
SFC a.c. 50µA,,d.c. 50µA
animal leakage current under single fault condition with mains voltage on the applied part,Limit: 50
uA (CF Type
)
Only qualified person who has adequate training, knowledge can perform these tests and make
record; If the equipment has malfunction during the test, it should be repaired.
1
)
Main Unit
Make sure the instrument is powered off.
Clean the instrument and accessories. Put on the dustproof cover after use.
Place the instrument in a dry and cool environment. Prevent from shaking violently when moving it
to another place.
Prevent any liquid from seeping into the ECG as it will affect its performance and safety.
Handle over the medical instrument to maintenance departments to check the ECG instrument
regularly.
2
)
Animal Cable
1
)
Check the animal cable is well-collected according to the following table. The resistor between
the electrode plug and plug pin of the animal cable plug should be less than 10Ω;
Note: the resistance or of animal cable with the function of defibrillating protection is around
10KΩ.
Wire plug symbol
R
L
F
RF
Plug pin location
9
10
11
14
2
)
Integrity of animal cable
,
including main cable and lead wires
,
should be checked regularly.
3
)
Knotting or crooking in closed angle will shorten the animal cable’s life, so please align the
animal cable and then collect the electrodes
4
)
Do not drag or twist the animal cable with excessive stress while using. Hold the connector
plugs instead of the cable when connect or disconnect the animal cable
5
)
Storing the lead wires in bigger wheel or swinging to avoid twisting or folding
6
)
Once damaged or aged animal cable has been found, replace it with new one immediately
7
)
Electrodes
Electrodes must be cleaned after use and be sure there is no remaining gel on them