53
Appendix C Specification
C.1 Technical Index
1) Main Machine
Lead
Standard 6 Leads
Sampling Method
6 leads simultaneously
Measurement Parameter
HR, PR interval, QRS duration, QT/QTC, R/QRS/T electric
axis, RV5/SVI amplitude
Input Method
Floating Ground input with defibrillator protection
Recording Mode
Manual Mode (Manual 1, Manual 2, Manual 3)
Rhythm Mode
Auto Mode
(
3CH
×
2+1R
)
Filter
Baseline Drift Filter: OFF/0.05Hz/ 0.16Hz/ 0.25Hz/ 0.32Hz/ 0.5Hz/ 0.67Hz;
EMG Filter: OFF/ 45Hz/40Hz/ 35Hz/30 Hz/ 25 Hz/20Hz
AC Filter: OFF /50Hz/ 60Hz
Low Pass Filter: OFF/75 Hz/ 100 Hz/ 150 Hz
Input Circuit Current
≤0.1 μA
Animal Leakage Currency
<
10μA
Time Constant
≥3.2 s
Frequency Response
0.05
~
150 Hz (-3 dB)
Noise Level
≤15 μVp-p
Paper Speed
5mm/s, 6.25 mm/s, 10mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
,
Tolerance ≤ ±5%
Standard Sensitivity
10 mm/mV±0.2 mm/mV
Sensitivity
2.5 mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV, Tolerance
±5%
Internal calibration voltage
1 mV
土
5%
Polarization Voltage
±300mV
CMRR
≥100dB