6.3 Setting the operational mode
This section describes how to set the operational mode. In the normal state, the
LCD controller displays pump status. To display the operational mode setting
screen, press the key “PRG.” for three seconds or longer. Pressing the key for
one second or longer again returns to the pump status display screen. Table 6.3
below shows indications and the details of the operational mode setting.
Table 6.3 Operational mode setting screen
Item
Indication
Description
Setting the pump operation
control mode
SET
CONTROL MODE?
Switches the control modes:
local ,remote, communication.
Setting the DIP switch
SET
DIP SW?
Performs the DIP switch settings
(see 6.4).
Setting the pump running
mode
SET
RUNNING MODE?
Switches the running modes:
NORMAL and S. ENERGY.
Setting the rotational speed
in the NOMAL mode
SET BP
NOMAL SPEED?
Sets the pump rotational speed
in the NOMAL mode.
Setting the rotational speed
in the LOW SPEED mode
SET BP
LOW SPEED?
Sets the pump rotational speed
in the LOW SPEED mode.
Setting the pump N2
warning
SET POINT
N2 FLOW LOW?
Sets the WARNING value for
N2 flow
Setting the WARNING value
for the back pressure
(option)
SET POINT
BACK PRESSURE?
Sets the WARNING value for
the back pressure.
Keys work as below for the setting screen.
START
: Valid
STOP
: Stops the pump.
RESET
: Resets WARNING and /or ALARM.
BZ.OFF
: Switches the DIP switch No.
△
: Sets the DIP switch to ON. Switches the display of the operational
mode setting screen.
▽
: Sets the DIP switch to OFF. Switches the display of the
operational mode setting screen.
ENTER
: Determines the selected setting.
See Fig. 6.3 for how to set the operational modes.