User manual SEFELEC 5x Series
27
v1.06
Ground continuity 32 A AC
(SEFELEC 32-C; SEFELEC 56/506-S)
Measurement current
Sinusoidal alternating with frequency of
50 Hz or 60 Hz depending on the mains
Adjustable in steps of
0.5 A
5.0 to 32.0 A AC
option 5.0 to 50.0 A ac
Generation and measurement accuracy
±(1% + 0.1 A)
option
±(1% + 0.2 A)
Maximum open circuit voltage
32 A generator 6 VAC ±10%
50 A generator 8 VAC ±10%
Voltage drop measurement
Measurement scope
0.01 to 9.99 V
Resolution
0.01 V
Accuracy (% of reading)
±(1.5% + 0.03 V)
Ground continuity measurement
This function calculates the value of the resistive circuit from the current generated and the voltage at
the terminals of the sample
Measurement scope
0.1 to 1,
000.0 mΩ
Resolution
0.1 kΩ
Indication of the unit
m
Ω = 0.001 Ω
Accuracy (% of the reading)
±(1.5% + 0.5 mΩ)
Resistance measurement range
According to the graph below: