The dielectric strength tests are performed at voltages adjustable from
100 VAC
to
5,000 VAC
and
100 VDC
to
6,000 VDC.
The rated current is
10 mA
for the
50 VA
models and
100 mA
for
500 VA models
. Breakdown voltages and currents are stored after circuit break and cut-off of the
high voltage.
The insulation resistance is measured at voltages of
-20 VDC
to
-1,000 VDC.
The device gives, in direct readings with display of the units, resistance values of
100 kΩ
to
200 GΩ
(
2
TΩ
on option).
The ground continuity resistance is measured at currents of
5 to 32 A AC
(50 A AC on option)
with open circuit voltages of 6 or
8 V
. The device gives, in direct readings with display of the
units, resistance values of 1
mΩ to 1
,
000 mΩ,
The devices in the
SEFELEC 5x
range are equipped as standard to communicate with any sys-
tem provided with an API,
RS232
,
Ethernet
or even
IEEE 488 (optional) interface.
OPTIONS (TO BE ORDERED ACCORDING TO THE REQUIREMENT)
Reference
Description
SEFO-IEEE488
IEEE 488-2 interface option
SEFO-CO160
red/green lamp for presence of high voltage option
SEFO-5X50A
50 A AC for ground continuity option
SEFO-5XREAR
rear outputs option
SEFO-5X3MY
short-circuit current limit of < 3 mA AC and DC in HV 50VA option
SEFO-5x2TB
2 TOhm in insulation option
SEFO-5XRC
remote control for accessories option