User manual SEFELEC 5x Series
101
v1.06
The lower part of the screen displays the different phases activated (1/n, 2/n, …) as well as a
zone of "Global" settings. Browsing in the different phases and in the Global mode is done either
by pressing at the bottom of the screen on the desired phase or by pressing the screen and
dragging the finger from right to left or left to right.
The "Global" mode allows to define the settings that will be applied to all the test phases ena-
bled, for example the AC or DC nature of voltage, modes and thresholds of cut-out as well as
the duration. The input of a setting is then applied to all phases, but it is possible to enter each
phase to change the settings.
To delete or add a phase (maximum = 7), select the phase concerned and press the Delete or to
Add icon:
Refer to paragraphs 9-1 to 9-6 of this manual for the adjust of the different settings.