Charging the Battery
The EasyStand Bantam Pow’r Up Lift uses two 12 volt sealed lead batteries rated at 2.9 amps each. An
EasyStand Bantam can be lifted approximately 100 times on one full battery charge. An audible alarm will
sound when the battery is low. Please charge the battery after the alarm sounds. A battery needing a full
recharge will take approximately 8 hours. The EasyStand Bantam can be raised to the standing position while
being charged. The function of the charger is to detect a full battery and then provide a constant potential
power supply to hold the battery at a full condition without overcharging. When the battery EMF rises to a fac-
tory preset point, the charger circuit will not allow the peak output voltage to exceed that value. This factory
setting protects a battery from excessive depletion of electrolyte, which can
occur from overcharging. The charger can be found in the tool pouch on the
back of the unit.
CAUTION:
Use only the Linak charger, model numbers 00CH01-XX OR
10CH01S-00, provided by Altimate Medical with your Pow’r Up option.
1. Plug the receptacle into a wall outlet. The light on the charger will be green
indicating an open circuit.
2.
FIG. A-
Plug the charger cord into the side of the hand control. The light on
the charger will turn yellow indicating the batteries are being charged.
3.
FIG. B
-The unit will arrive unplugged from the battery, you need to plug in
the white cord that supplies power to the actuator and the black cord that
supplies power to the remote. The black cord has a notch on it to orient the
connector. Fully charge battery before use.
Note:
The white cord has a rubber washer on it, which makes it very sticky and
sometimes will not get pushed in completely, make sure to push the cord all
the way into the battery.
4. When the batteries are charged, the light on the charger will turn green.
5. Fully charge the batteries every three months or less during periods of
intermittent or non-use.
CAUTION:
The Bantam Pow’r Up Option is not designed for continuous use.
Duty Cycle 5% Max, 1 min. / 19 min.
Raising the Lift -
Press the UP (arrow up) button on the hand control to lift the
seat.
Lowering the Lift -
Press the DOWN (arrow down) button on the hand control
to lower the seat.
CAUTION:
Ensure the cables are attached properly per the instructions for use. Do not allow the cables to
drag on the floor or to catch on objects.
Pow’r Up Lift Option Usage
FIG. A
FIG. B
Pow’r Up Lift Option
Page 19
!
!
!
The hand pendant
is a Type BF applied
part.