PROCESSOR Voltage
9B8948 – 15JUN98
9
[2] Remove the ELECTRICAL COVER.
[3] Determine the voltage of the power source at the site. It must be 220, 230 or 240 V AC.
[4] Locate TERMINAL BLOCK TB51 inside the PROCESSOR.
[5] Use the following table to select the correct voltage setup for the PROCESSOR.
[6] Move WIRE 6 to the correct location.
[7] Install the ELECTRICAL COVER.
[8] Close the PROCESSOR and install the following:
• DEVELOPER and FIXER DRAIN VALVES and BRACKET
• WASH WATER OVERFLOW and WASH DRAIN FITTINGS
• 2 SCREWS on the top of the PROCESSOR
Note
Be careful not to bend the HOSES when you close the PROCESSOR.
[9] Label the DATA PLATE on the back of the PROCESSOR with the internal voltage setup.
[10] Advance to Installing the PROCESSOR on
.
Voltage at site
TERMINAL on TB51
220 V AC
5
230 V AC
6
240 V AC
7
H164_0077HC
1
2
3
4
5
6
7
5, 6, and 7
TERMINALS
SCREW
WIRE 6
ELECTRICAL COVER
H164_0077HCA