Step 9:
Now return to the TOP.SET page by clicking on the TO SET shortcut key in the top
right corner. Now press the FIELD CALd Bu on.
Step 10:
You should now be on the SET.CALd page. The newly calculated Dynamic Calibra on
can be seen on the le side of the page, while the current Dynamic calibra on can be
seen on the right side of the screen. If you would like to change your current se ngs
with the se ngs that you calculated please move on to Step 11. If you would like to
keep your current se ngs, please press the CANCEL Bu on.
Step 11:
To accept the new Dynamic Calibra on press the YELLOW CACULATED to CURRENT
Bu on and press the OK Bu on. A pop up screen will appear, asking if you are sure
you would like to accept this value. Press YES.