The
User
Settings
Log
indicates
each
time
that
the
user
settings
have
been
changed
and
what
the
new
settings
are.
A
typical
Userset
page
can
be
seen
below.
The
MTR
Setup
Sheet
allows
data
to
be
loaded
to
it
and
to
calculate
the
Standard
Deviation
and
the
Percent
Error
of
the
Electronics’
readings.
This
file
is
stores
site
calibration
data
to
the
compact
flash
from
the
SET.SAVE
page.
A
standard
MTR
Setup
Sheet
can
be
seen
below.
View
Data
This
option
is
not
used.
DATE
TIME
EVENT
ZERO
CALs
CALd
RANGE
INHIBIT
FILTER
CWratio
Full
Scale
6/2/2009 14:47:54 Power
On
7FEF
10D0
2710
4
2
14
180
10000
6/2/2009 14:49:08
zero
7FDC
10D0
2710
4
2
14
180
10000
6/2/2009 14:49:30
CALs
7FDC
10F2
2710
4
2
14
180
10000
6/2/2009 14:55:11 Power
On
7FDC
10F2
2710
4
2
14
180
10000
MM
‐
DD
‐
YY:
5/21/2009
TIME:
15:11:47
Material:
rice
shortcb
Technician:
sjt
Location:
wh
Type:
Type
I
Reverse
Model:
CFM
‐
12
Feed:
Belt
Full
Scale:
1000
Zero:
7F5C
CALs:
0C5D
CALd:
3198
RUN
ACTIVE
METER
WEIGHT ACTUAL
WEIGHT ACTUAL/METER %ERROR
1
1
30.267
30.46
1.0064
‐
0.63
2
1
30.567
30.72
1.005
‐
0.5
3
1
26.433
26.56
1.0048
‐
0.48
4
1
30.567
30.72
1.005
‐
0.5
5
1
28.233
28.42
1.0066
‐
0.66
Average:
1.0056
STD:
0.0009
%STD:
0.0857