Step
3:
After
product
has
stopped
flowing
through
the
meter,
go
to
the
SET.FIELD
page
located
under
the
SET
option
on
the
MAIN
MENU.
HINT:
Press
the
MENU
softkey,
then
press
SET,
and
finally,
FIELD
DATA.
Step
4:
Click
on
the
“ENTER
RUN”
Button
and
you
should
see
the
GREEN
value
(CF
Meter)
change
from
“0.000”
to
a
number
signifying
the
weight
that
the
electronics
calculates
passed
through
the
meter.
Step
5:
Weigh
the
collected
material
on
a
static
scale
and
record
the
weight
in
the
YELLOW
block
(Actual)
next
to
the
value
automatically
entered
in
STEP
4.
Press
the
YELLOW
box
to
bring
up
a
numeric
keypad.
After
entering
the
value,
press
the
NEXT
RUN
Button
to
prepare
for
the
next
run.
It
is
important
that
the
container
holding
the
material
not
be
included
in
the
weight.
The
only
weight
that
is
important
is
the
material
that
actually
went
through
the
meter.
Step
6:
Return
to
the
TOP.RUN
page
and
press
the
RED
RESET
Button
on
the
bottom
left.
Step
7:
Repeat
Steps
2
‐
6
four
more
times
to
complete
the
5
Field
Data
runs.
Step
8:
Once
all
five
runs
are
completed
click
on
the
RUN
Number
next
to
each
set
of
data
in
order
to
select
each
data
RUN.
When
selected
the
circles
next
to
the
numbers
will
become
green
as
seen
in
the
picture
above.