Save Parameter:
After performing any of the above calibrations select
[Param save]
to save the calibration
setting.
(Password: 9577)
Function Settings
The function menu is used to change Alarm, Parallel Load, Prepare Time, Restore Time,
Company, DAC Type, and String Group settings.
Function Menu
Description of each function:
[Alarm]:
Notifies user when discharge test is complete using a BEEP sound
[Parallel]:
Allows for discharge using two SLB load banks in parallel.
[Prepare Time]:
When set, a minute of prepare time is given before any discharge test
performed.
[Restore Time]:
When set, a minute of restore time is given after completing a discharge.
[Company]:
Allows you to specify company information for any discharge test.
[DAC Type]:
Sets DAC cell voltage type.
NOTE:
For all standard DACs, this should always be
set to [1.2, 2, 6, 12V] in order to have proper communication to the DACs.
[Str Grouped]:
Used to set parallel strings using DAC. These DACs are grouped to fit
different strings in the way of String 1 (DAC 1, 2…), String 2 (DAC1, 2…) and so forth.
For detailed discharge instructions please refer to the parallel testing section
Service & Maintenance
Cleaning
To clean Eagle Eye SLB-Series load banks simply wipe the units down with a damp cloth and
mild soap. Do not use abrasives, solvents, or alcohol as they can deform and or discolor the
unit. Occasionally, the load bank and its components should be blown off with air to remove
dust buildup from the mosfets and control boards. Do not apply extreme psi directly onto the
boards when cleaning.
NOTE
:
Be sure to unplug all cables before cleaning the unit.
Storage
Store all Eagle Eye SLB-Series units in a cool, well ventilated place with relatively low
humidity. If the unit came with a case, store it in the case whenever the unit is not in use.