Time and Date Setting
Setting the system time and date correctly should be done before any discharge tests are
performed.
To set system time:
1. Select
[Syst time]
under the
[System manage]
menu
2. Use the
up/down
arrow keys to change number orientation
3. Use the
left/right
arrow keys to change number values
4. Press the
ENT
button to save the time/date setting
Parameter Management
Parameter management is used for various calibration and parameter settings.
Param manage
Zero calib
Measure calib
Mainboard Para
Parameter Management Menu
Zero Calibration
To ensure measurement accuracy, zero calibration should be performed when the load bank
was unable to collect discharge voltage and current in normal operation.
To perform zero calibration:
1. Disconnect all measurement cables from the load bank and batteries
2. Select
[Zero calib]
under the
[Param manage]
menu
3. Press the
ENT
button to start calibration
Time setting
08/30/201
6
12:30:00