15.1.2 Exchanging the Pump Lever
1. Dismount the pump as described in section 14.3.
2. The pump lever is fixed with 2 pins and can be dismounted carefully by
hand. Do not use any tools.
3. Mount the new pump lever by shifting it in place.
15.2
Firmware Update
Your
Eagle Eye
representative will inform you when a new firmware update
for your
SG-Ultra Max
is available. After receiving the update file, you can
import it into the instrument via Bluetooth. For details about setting up a
Bluetooth connection, see section
1.
Save the update file (extension *.afp) on a PC (e.g. on the desktop).
2.
On the PC, right-click the update file and select "Send it to Bluetooth".
3.
On
SG-Ultra Max
tap <Menu> and select
Setup
> Data Transfer
>
Firmware
Update
.
4.
Tap <Start> to start the automatic update procedure.
15.3 System Information
The instrument holds system information comprising:
- Instrument type
- Serial number of the instrument
- Firmware version
- Hardware version
- Serial number of the density measuring cell
- Deviation density adjustment
- Printer name
- Printer address
- PC name
- PC address
- Number of performed boot ups
- Operating time
Accessing system information
1. Tap <Menu> and select
Service > System Information
.
2. Tap <Back> repeatedly to return to the main screen.