35
EN
6. After the procedure has finished, the density deviation to the previous
adjustment is displayed.
7. To save the adjustment, tap <Save>.
To refuse the adjustment, tap <Cancel>.
10.3 Performing a Custom Adjustment
You can use any other reference liquid with a known density at 20 °C and
known temperature coefficient Alpha for a custom adjustment.
We recommend using the custom adjustment e.g. if you measure only liquids
in a specific density range (e.g. high density) and want to achieve more
accurate results in this range.
Performing a custom adjustment
1. Clean the measuring cell as described in section 14.
2. Rinse the measuring cell until no more cleaning agent residues are
present.
3. Tap <Menu> and select
Adjustments > Custom Adjustment
.
4. Fill reference liquid (15 °C to 25 °C) into the instrument.
5. Tap <Start>.
6. Tap <Edit> and enter the density of the reference liquid at the stated
reference temperature.
7. Tap <Next>.
8. Tap <Edit> and enter the temperature coefficient Alpha of the reference
liquid.
9. Tap <Next>.
10.After the procedure has finished, the density deviation to the previous
adjustment is displayed.
11.To save the adjustment, tap <Save>.
To refuse the adjustment, tap <Cancel>.