C H A P T E R 2
Load Cell Calibration
Torque Cell Installation Guide for Model 250i/250iP DynoWare RT Dynamometers
2-2
LOAD CELL CALIBRATION
This section provides instructions for calibrating the load cell. Follow the directions on
the screen exactly. Failure to perform the directions accurately will result in improper
torque values.
You will need the following parts:
1
Verify the dyno is connected to power. Refer to “Main Dyno Power” on page 1-3.
2
Double-click the Power Core program icon.
3
Click
Dyno Control
from the Application Launcher.
4
Verify you are connected to the DynoWare RT main module.
Note:
For more information on connecting to the dyno electronics, refer to the
Power Core Quick Start Guide (on your Power Core CD or at www.dynojet.com) or
the Power Core Online Help.
5
From the Configuration ribbon, click
Load Cell Calibration
.
6
Remove the calibration arm and mass if it is installed.
7
Release the dyno brake. Make sure the drums are free to rotate.
Note:
There should not be anything resting on the eddy current brake or the dyno
drum during this procedure.
8
Click
Next
to zero the connected torque cell.
Figure 2-1: Zero Calibration Window
part
description
part
description
weight, 25 pounds (4)
P/N 35430899
calibration arm assembly
P/N 61319001