2 MediTEST User Guide Page 12
-
4
SYSTEM CALIBRATION
12.5 SYSTEM LEAK TEST
It is possible to conduct a system test chamber leak test (located in the
SYSTEM SETTINGS
screen.
This may be of use if erroneous test results suggest a system leak has arisen, such as due to a worn
drawer seal.
Before performing a System Leak Test, ensure valid test parameters are entered (similar to Fig 12.3).
These should be as listed in section 5.3. If they are not, or there is no documentation relating to their
current setting, then contact Dynascan for advice.
Figure 12.5
To perform a System Leak Test:
Ensure no packs are inserted into the test chamber and close the Drawer.
Enter the
SYSTEM SETTING
screen and press
SYSTEM LEAK TEST (
Fig 12.6)
Press
START
. The leak test cycle will execute.
At the end of the test a PASS or FAIL result will be posted (as determined by the test parameters).
Pressing
OK
will save a record. Pressing
CANCEL
will exit.
Figure 12.6