-oL-
〗
〖
End
〗
Calibration finish
〖
3000
〗
Attention: the standard load should be more than
20% capacity, it’s better to
use full capacity standard load
CHAPTER 6
DISPLAY ILLUSTRATION
DISPLAY
ILLUSTRATION
REMARK
〖
noSIg
〗
Without wireless signal
Distance too far.
〖
SEtUP
〗
Enter into parameter setting
〖
UAdJ
〗
Enter voltage calibration
〖
LoAd
〗
Calibration point
〖
¯ ¯ ¯ ¯ ¯
〗
Exceed high limit
Tare weight can
’t exceed full capacity
〖﹍﹍
〗
Exceed low limit
Tare weight can
’t be negative
〖
-----
〗
Waiting stable
〖
Err10
〗
Weight less than Min. Capacity
Can
’t accumulate the value
〖
Err11
〗
Accumulated times overflow
Can
’t accumulate after 30times
〖
Err12
〗
Accumulated weight overflow
Can
’t accumulate after 99999
〖
Err13
〗
Error in repeat accumulation
Can
’t accumulate one weight repeatedly
〖
no***
〗
Current accumulation times
〖
H****
〗
Front four digit of accumulated
Total weight=front four digit + rare four digit
〖
L****
〗
Rare four digit of accumulated
Total weight=front four digit + rare four digit
〖
CLr
〗
Ask if you really want to delete
accumulated weight
In case error deletion
〖
noCLr
〗
Give up deletion
〖
88888
〗
Confirm deletion
〖
¯ ¯ ¯
〗
Input value is too large
When you input tare or weight value
〖
﹍
〗
Input value is too small
When you input tare or weight value
〖
noACC
〗
No any accumulated content
when you check accumulation
〖
-oL0-
〗
Overload warning
Tare + Net weight exceed full ca 9e
〖
-oL1-
〗
Overload warning
Tare + Net weight exceed full capacity 125%
〖
-Lb-
〗
Low battery warning
Turn off automatically one minute later
〖
U*.**
〗
The voltage of current battery
〖
End
〗
End
when parameter setting or calibration ready
〖
OFF
〗
Turn off
〖
Unstb
〗
Input value before STB light on