background image

                                                                                                    

 

d

MODE

 

Used as function shift, or be combined with other keys. 

 Press

MODE

key once, it display 

“OFF”, press

ENTER

Key , it will wireless turn off 

dynamometer, and indicator display 

“noSIG”, means no signal. 

 Press

MODE

key twice, it display 

“SETUP”, Press 

ENTER

Key access Parameter 

Setting status. After setting finish, press 

MODE

key  to confirm setting and back to 

test mode. 

 Press

MODE

key 3   times, it   display 

“dC”, Press

ENTER

Key   to read current   battery 

voltage, press

MODE

key again back to test mode. 

  Press 

MODE

key  4  times,  it  display 

“Unit”,  Press 

ENTER

Key  access  unit 

selection. Press

ZERO

key to select unit, press

MODE

key again back to test mode. 

  Press

MODE

key  5  times,  i t   display 

“ACCU”,  Press 

ENTER

Key  access 

Accumulation status, press

MODE

key again back to test mode. 

  Press

MODE

 key  6  times,  i t   display 

“SACCU”, Press

ENTER

 Key   

access Accumulation Search status, press

MODE

key again back to test mode. 

 Press

MODE

 key 7 times, display 

“CLEAr”, Press

ENTER

 Key access Accumulation clear 

status, display 

“CLr    1”, Press

TARE

Key, dis

play“noCLr”. press

MODE

key again back to 

test mode. 

 Press

MODE

  key 8 times, display 

“CLIbr”, Press

ENTER

Key enter into calibration 

program, after calibration finish, press

MENU

key back to test MENU. Detail see Pg.13 

 Press

MODE

  key 9 times, it display 

“ESC”, Press 

ENTER

Key and It will back to test mode. 

 

e

ENTER

 

Used as function selection, normally combined with 

MODE

Key 

f. 

PEAK

 

When load is changing, press this key will catch and display the maximum reading of the load.

 

Press this key again, reading will return to normal. 

 

 

CHAPTER 4    DYNAMOMETER FUNCTION GUIDE 

 

 

Note

(    ) means the   key on the dynamometer   

 

means the key on the indicator 

 

means the display content   

1. ON/OFF 

TURN ON dynamometer 

OPERATION

 

DISPLAY

 

ILLUSTRATION

 

(ON/OFF)

 

88888

 

Display twice, self test

 

 

  Ert   

 

Display twice

 

 

u=3.72

 

Display current software version

 

 

CH=E3

 

Display current wireless channel E3

 

 

  5000

 

  Display capacity, e.g. 5t

 

 

U=3.98

 

Display current battery voltage is 3.98

 

 

–––––

 

Waiting stable 

 

        0

 

Display current load, generally is 0 

                                                                                                                                                                     

USER MANUAL                                                                7                                                                 

Summary of Contents for W3-01

Page 1: ...WIRELESS DYNAMOMETER USER MANUAL Model W3 USER MANUAL ...

Page 2: ...RELESS INDICATOR OPERATION GUIDE 6 CHAPTER 4 DYMOMETER FUNCTION GUIDE 7 1 On off 7 2 Zero 8 3 Tare 8 4 Peak Hold 8 5 Accumulation 9 6 Accumulation Search 9 7 Accumulation Clear 9 8 Battery Voltage 9 9 Unit 9 10 Return 10 11 Set point 10 CHAPTER 5 PARAMETER SETTING CALIBRATION 11 1 Key function 11 2 Parameter setting 11 3 Calibration 13 CHAPTER 6 SIGNAL ILLUSTRATION 14 CHAPTER 7 TROUBLESHOOTING GUI...

Page 3: ...hich may cause incorrect reading 5 Do not attempt to open this Dynamometer no user serviceable parts inside 6 Do not remove wire rope stopper from hook For safety reasons always apply it 7 Remove all loads from shackle or hook when not in used 8 Before weighing check and make sure that all hanging load receiving elements devices are in good condition 9 Check hook shackle safety pins and latches pe...

Page 4: ... for limit weighing The dynamometer is powered by 3 standard LR6 AA size alkaline batteries All commonly used internationally recognized units are available kilograms kg short Tons t pounds lb Newton and kilo newton kN Operation through wireless handheld indicator easier to calibrate with password 4 local mechanical keys ON OFF ZERO PEAK and Unit Change Low battery warning RF wireless handheld ind...

Page 5: ...IMENSIONS Dimensions shown are nominal and subject to tolerances MODEL CAP A mm B mm C mm D mm φ mm H mm Material W3 1 1t 245 112 37 190 43 335 Aluminum W3 2 2t 260 123 37 195 51 365 Aluminum W3 3 3t 260 123 37 195 51 365 Aluminum W3 5 5t 285 123 57 210 58 405 Aluminum W3 10 10t 320 120 57 230 92 535 Alloy Steel W3 20 20t 375 128 74 260 127 660 Alloy Steel W3 30 30 420 138 82 280 146 740 Alloy Ste...

Page 6: ...can be used for about 80hours under continuous working Low battery warning When the digits flashes it means the battery needs to be replaced The Dynamometer or wireless indicator will power off automatically after one hour without operation CHAPTER 2 DISPLAY AND KEYS 1 LCD DISPLAY a 25mm 1 5digits LCD with all functions and units b Wireless signal it Appear means signal stable it disappear while s...

Page 7: ... to normal CHAPTER 3 WIRELESS INDICATOR OPERATION GUIDE a ON OFF 1 Press ON OFF Key 1 Second the indicator turn on 2 In ON status Press ON OFF Key 1 Second the indicator turn off b ZERO 1 In testing status there are small digits on screen press this key to obtain the zero reading 2 In Parameter Setting status it used as scrolling up digit c TARE 1 When there is tare weight such as container after ...

Page 8: ... key 7 times display CLEAr Press ENTER Key access Accumulation clear status display CLr 1 Press TARE Key display noCLr press MODE key again back to test mode Press MODE key 8 times display CLIbr Press ENTER Key enter into calibration program after calibration finish press MENU key back to test MENU Detail see Pg 13 Press MODE key 9 times it display ESC Press ENTER Key and It will back to test mode...

Page 9: ...ting the dynamometer will display after no operation at 15minutes to save power and it will turn off automatically after 60minutes TURN OFF indicator 1 Manual Power OFF Press ON OFF Key 1 Second 2 Auto Power OFF The indicator will turn off automatically after 3 minutes if no operation 2 ZERO OPERATION DISPLAY When Dynamometer turn on generally it displays 0 if display small digits when no load pre...

Page 10: ...N CLEAR OPERATION DISPLAY ILLUSTRATION Press MODE key 7times CLEAr Press ENTER CLr 1 Ask if you want to clear or not noCLr Press MODE key Cancel clear 88888 Press ENTER keyConfirm clear 8 BATTERY VOLTAGE OPERATION DISPLAY ILLUSTRATION Press MODE key 3 times Display dC press ENTER key U Display current battery voltage Press MODE key again Return Note Battery voltage normally between U 3 50 to U 4 7...

Page 11: ...or will alarm when load smaller than the value Press ENTER to confirm 02000 New dynamometer it display 02000 if parameter set the screen display set value Press ZERO and TARE to change value 01000 Set 1 HI or 1 LO Value e g 1000kg Press ENTER to confirm SP2 Set point 2 Press ENTER to confirm 2 OFF New dynamometer it display 2 OFF if parameter set it display set value Press ZERO to choose 2 HI or 2...

Page 12: ...irm Id 02 New dynamometer the screen display Id 02 if dynamometer calibrated it will display calibrated division Press ZERO to choose Id 02 Choose division it display 01 02 05 10 20circularly 5000kg dynamometer division is 2kg so choose Id 02 Press ENTER Pt 0 New dynamometer it display Pt 0 if dynamometer calibrated it will display calibrated decimal position Press ZERO to choose Pt 0 Choose decim...

Page 13: ...hoose turn off method 0 2 total 3 options 0 turn off by hand 1 if no operation the screen display after 15minutes to save power 2 if no operation the screen display after 15minutes and turn off automatically after 60minutes Generally choose OFF 2 when ex stock Press ENTER SP1 Set point 1 Press ENTER to confirm 1 OFF New dynamometer it display 1 OFF if parameter set it display set value Press ZERO ...

Page 14: ... 5kg 10 05 0 15000kg 5kg 15 05 0 20000kg 10kg 20 10 0 30000kg 10kg 30 10 0 50000kg 20kg 50 20 0 100t 200t use specail software no need to set FS Id Pt 3 Calibration User must set all parameter before calibration OPERATION DISPLAY ILLUSTRATION Turn on dynamometer and indicator 0 Take Cap 5000kg dynamometer as example make sure it display 0 before calibration you can hang tare weight first then turn...

Page 15: ...ion Can t accumulate one weight repeatedly no Current accumulation times H Front four digit of accumulated Total weight front four digit rare four digit L Rare four digit of accumulated Total weight front four digit rare four digit CLr Ask if you really want to delete accumulated weight In case error deletion noCLr Give up deletion 88888 Confirm deletion Input value is too large When you input tar...

Page 16: ...n After turning on heating 3 5 minutes Displayed weight shows large error Dynamometer not Zeroed before applying weight Depress ZERO before applying weight Requires recalibration See calibration Kg lb wrong selection See operation Wireless distance shortened Wireless indicator s battery is low Replace battery CHAPTER 8 RS232 PROTOCOL Pre code Data ASCII 0X 39 39 F0 F0 FF AA data H data data data d...