BlueWave® MX-275 Emitter User Guide
14
Product Cleaning and Care
Product cleaning is limited to wiping the product with a damp cloth. Do not soak.
Isopropanol Alcohol or household cleaners may be used for cleaning the product.
Always inspect the quartz window for cleanliness before use. Foreign material can
cause permanent damage to the window. Clean with Isopropanol Alcohol to remove
smudges or foreign material. Damaged or permanently etched windows should be
replaced.
Do not use compressed air to removed particle debris inside the emitter as it may
damage the high-speed cooling fan.
Glass Window Replacement
If the glass window in your BlueWave MX-275 emitter becomes broken or contaminated
and needs to be replaced, Dymax offers a glass window replacement kit (PN 43362).
Follow the instructions below for proper replacement.
1.
Before starting, disconnect power to the unit.
2.
Lay the emitter on its side with the cover facing forward. Using a Philips #1
screwdriver, loosen and fully remove the two screws (Figure 4).
3.
Remove the glass retaining plate being careful not to lose the O-rings.
Figure 4. Remove Screws