BlueWave® MX-275 Emitter User Guide
13
Troubleshooting & Maintenance
Problem
Possible Cause
Corrective Action
BlueWave® MX-275
LED does not
produce light
LED intensity adjustment
set to 0% or too low
Increase LED intensity setting.
LED cycle time is set to 0
seconds
0 Seconds sets Manual Mode and
requires a trigger
Interlock is open
Verify interlock jumpers are in
place.
Verify PLC command structure for
PLC mode
Interface cable connections
loose or damaged
Check connections and condition
of interface cable
Trigger setting not
matched to input.
Trigger setting on Admin screen
should match the desired input
trigger channel
LED head is not connected
to the correct port/channel
Verify that the head is connected
to the desired port/channel
BlueWave® MX-275
LED suddenly stops
producing light
Over-temperature
shutdown was triggered
Verify alarms
Footswitch defective
Activate unit using the front
control panel. Replace the
footswitch if the unit operates
from the front control panel
Interlock is open
Verify interlock jumpers are in
place
Verify PLC command structure for
PLC mode
BlueWave® MX-275
LED provides only
low-intensity light
LED intensity adjustment
set to minimum
Increase LED intensity setting on
Admin settings or I/O input for PLC
mode
Contaminated/dirty lens
optics
Clean the surface of the lens