Troubleshooting
WARNING!
Only qualified maintenance personnel should attempt the following procedures:
Table 4.
Troubleshooting Chart for BlueWave AX-550 Flood Curing System
Problem
Possible Cause
Corrective Action
BlueWave AX-550 system
does not power up
Power cord not plugged
in or damaged.
Check power connection and condition at power
source and AC Inlet fuse of the unit.
No electrical power at
Receptacle.
Test Receptacle for power.
BlueWave AX-550 system
powers up but the emitter
does not emit light
MASTER INTERLOCK and/or
LED INHIBIT is open.
Verify both of MASTER INTERLOCK and LED
INHIBIT are connected to common directly or
through the safety sensor.
Verify PLC command structure for PLC mode.
Emitter is not connected to the
Controller.
Verify that the Emitter is connected to the
Controller.
BlueWave AX-550 system is
operating normally and the
emitter suddenly stops
emitting light
Over-temperature shutdown was
triggered.
Verify error information.
MASTER INTERLOCK or LED
INHIBIT is open.
Verify interlock jumpers are in place.
Verify PLC command structure for PLC mode.
LED emitter provides only
low-intensity light
LED intensity set too low.
Increase LED intensity setting on Admin Settings
or I/O input for PLC Mode.
Contaminated/dirty lens optics.
Clean the surface of the Lens.
Footswitch does not function
correctly
Footswitch not connected or
connected to wrong I/O
connector pins.
Verify connections to Pins 6 & COM of the I/O
connector.
Footswitch defective.
Activate unit using the front panel rotary
pushbutton. Replace the footswitch if the unit
operates from the rotary pushbutton.
Error Screen
If the unit is powered up, and some error happens, the error screen
will appear.
The error screen will display the system serial number, controller and
driver firmware versions, a technical support phone number for more
information, and the current error description.
Figure 54.
Error Screen