18
USE R GUI DE
|
BL UE WAVE
®
A X -5 5 0
Figure 26.
Padlock location
Figure 27.
Information screen
System Initializing
The system is designed to work in any one of three modes of operation as
indicated in the upper left of the display:
Admin Mode:
The administration mode allows full control of
system functions.
Production Mode:
This mode doesn’t allow changes to the
irradiation parameters.
PLC Mode:
in PLC mode, an external PLC can control the unit.*
*
NOTE:
PLC Mode icon is located in the upper right of the display.
To enter PLC mode, connect the external PLC system and toggle the
PLC
ENABLE (Pin 12)
input Low to 0VDC. The PLC icon will be seen on the
screen automatically. Leave the PLC ENABLE open to enter the
Administrator or Production mode.
Selecting the padlock button switches between Administration and
Production modes, a password will be required to enter admin mode.
The default password to enter admin mode is
00000
. To change the
password, please
see
“System Settings”
and Figure 32
NOTE
: Switching between PLC and non-PLC mode can happen when the
system is powered ON or OFF, but the LED must be OFF for safety.
Information
Information:
Displays system information.
Only available in Admin mode,
using the rotary pushbutton select the
information icon in the upper right area of the LCD screen.
Push the rotary pushbutton
and the “Information” screen will appear.
The following Controller/Emitter system information will be
displayed.
1.
The Controller Serial Number and Firmware Revision.
2.
The Emitter Array Type (RediCure, PrimeCure, VisiCure).
3.
The Emitter Array Serial Number and Version.
4.
The Calibration Date (yymm): target intensity.
5.
The maximum (temperature, current draw, voltage level),
experienced by the Emitter Array.
6.
LED HRS (Total time the emitter array has been illuminated).
7.
LED PRH (Power Rated Hours = Total time the emitter array
has been illuminated* Power Setting / 100).
Push the rotary pushbutton to return to the main screen.